Resonance R16M Mode D'emploi page 102

Analyseur d'oreille moyenne
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

The noise, progress and sweep indicators are updated in real time while the test is in progress.
Press SETUP (see picture beside) to open the test setup page.
The menu allows for the setting of a few test parameters:
Adjustable parameters:
Max Sweep
Max test time. By touching this field, the maximum test time, between 30 and 75 seconds, can be
selected. When a valid result is obtained the test is automatically stopped.
Artefact Level
In this field the maximum acceptable noise level can be selected. When the noise exceeds the set
value, the measurement is rejected (artefact). By touching this field, a value between 20 and 60 dB SPL
can be selected.
Stimulus Level
Allows for the setting of the stimulus level. By touching this field, a value between 70 and 90 dB peSPL
can be selected.
A maximum of 3 protocols of predefined values can be saved by keeping the numbered fields on the
right side of the screen pressed for a few seconds and until they turn green.
R16M User's Manual – Page 102