5.4 OAE Setup

Press the OAE tab to enter the respective setup page:
End of test criterion after obtaining the PASS
With this key (see adjacent image) it is possible to establish the criterion for ending the
test in case of Pass: it is possible to end the test just when the Pass is obtained or, even if
the pass was obtained, continue to the end of time and see overall result.
DP-OAE test frequency scan
Pressing this button (see adjacent image) changes the scanning order of the DP test
frequencies: from the minor to the major or vice versa.
OAE Probe check
This function allows you to check the operation of the OAE probe. Press the key to access
the verification menu, it is necessary to insert the probe in the test cavity provided to start
the verification.
The functionality will be checked and the result will be provided on the screen; if OK, the
tests can be carried out, otherwise the probe must be checked for: tip cleaning; tip
correctly inserter into probe body; front part probe coupling (both brackets hooked), see
chap. 11.1 for more information.
If the malfunction persists, contact technical support.
R16M User's Manual – Page 121