Télécharger Imprimer la page

Dräger X-am 1100 Notice D'utilisation page 113

Masquer les pouces Voir aussi pour X-am 1100:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 12
Notes on Approval
1 Marking
Type: LQG 00xx
Dräger Safety
23560 Lübeck
Germany
9N54
LR97594
C22.2 No. 152
I M2 / II 2G
0158
BVS 06 ATEX G 006X
PFG 06 G 001
Ex ia d I/IIC T4/T3
DEMKO 06 ATEX 140055X
Um=4.6V Im=1.3A
IECEx ULD 06.0001X
Ex ia s Zone 0 I/IIC TC T4/T3
ANZEx 06.3037X
Only as to intrinsic safety for use in haz. loc.
Intrinsically safe / Securite Intrinseque Ex ia
Class I&II, Div. 1 Gr. A,B,C,D,E,F,G TC T4/T3
CSA: Class I, Div. 1, Gr. A,B,C,D TC T4/T3
–20°C ≤ Ta ≤ +50/+40°C
TC & Ambient Temperature: see Battery Pack!
Warning: Read manual for safety precautions.
Do not change or charge batteries in haz. loc.
1)
Serial No.
––––––––––––––––
1) The year of manufacture is indicated by the third letter in the serial
number: T = 2003, U = 2004, W = 2005, X = 2006, Y = 2007, Z = 2008,
A = 2009, B = 2010, C = 2011, etc.
Example: Serial No. ARUH-0054: the third letter is U, which means that
the unit was manufactured in 2004.
PB ExdiaI X / 1ExdiaIICT4/T3 X
ГБ05
®
Battery Pack Type ABT 0000
Temperature Class T4
o
o
–20
C ≤ Ta ≤ +50
C
if used with alkaline batteries
— Energizer No. E91
— Energizer No. EN91 (industrial)
— Varta Type 4106 (power one)
Temperature Class T3
o
o
–20
C ≤ Ta ≤ +40
C
if used with NiMH batteries
— GP 180AAHC (1800 mAh)
Battery Pack Type HBT 0000
Temperature Class T4
o
o
–20
C ≤ Ta ≤ +50
C
2 Intended Operating Area and Operating
Conditions
Hazardous Areas classified by Zones
The device is intended to be used in hazardous areas or
mines susceptible to firedamp classified zone 1 or
zone 2, within a temperature range of –20
where gases of explosion groups IIA, IIB or IIC and tem-
perature class T3 or T4 (depending on battery pack and
batteries) may be present.
If used in mines, the device is only to be used in areas
known to have a low risk of mechanical impact.
Hazardous Areas classified by Divisions
The device is intended to be used in hazardous areas or
mines susceptible to firedamp classified Class I&II, Div. 1
or Div. 2, within a temperature range of –20
where gases or dusts of groups A, B, C, D or E, F, G and
temperature class T3 or T4 (depending on battery pack
and batteries) may be present.
3 Instructions for Safety
WARNING:
To reduce the risk of explosion, do not mix new
batteries with used batteries, or mix batteries
of different type or manufacturer.
WARNING:
Remove batteries before servicing.
WARNING:
Substitution of components may impair Intrin-
sic Safety.
CAUTION:
Not tested in oxygen enriched atmospheres
(>21% O
).
2
Only the combustible gas detection portion of this instru-
ment has been assessed for performance.
CAUTION:
high off scale readings may indicate an explo-
sive concentration.
CAUTION:
Before each day's usage sensitivity must be
tested on a known concentration of the gas to
be detected equivalent to 25 - 50% of full scale
concentration. Accuracy must be within 0 to
+20% of actual. Accuracy may be corrected by
calibration.
o
o
C...+50
C,
o
o
C...+50
C,
113

Publicité

loading

Ce manuel est également adapté pour:

X-am 1700X-am 2000