Table 4
Recommended safe distances between portable or mobile HF telecommunication devices and
the Suprasorb

CNP P3 therapy unit

®
The Suprasorb
CNP P3 therapy unit is intended for operation in an electromagnetic environment
®
as described below in which HF interference is controlled.
The customer or user of the Suprasorb
interference by ensuring maintenance of the minimum distance between portable and mobile
HF telecommunication devices (transmitters) and the Suprasorb
the output power of the communication device, as specified below.
Protective distance depending on transmitter frequency in m
Nominal capacity
150 kHz to 80 MHz
of transmitter
0.01
0.12
0.1
0.37
1
1.17
10
3.69
100
11.67
For transmitters with maximum nominal power not given in the table above, the recommended protective distance (d) in metres (m)
can be calculated using the formula in the respective column, where P is the transmitter's maximum nominal power in watts (W)
in accordance with the transmitter manufacturer's specifications.
NOTE 1: The higher frequency range applies at 80 MHz and 800 MHz.
NOTE 2: These guidelines may not be applicable in all cases. The transmission of electromagnetic parameters is influenced by
the extent to which they are absorbed and reflected by buildings, objects and people.
CNP P3 therapy unit can help avoid electromagnetic
®
3
5 ,
d
[
]
P
=
V
1
®
80 MHz to 800 MHz
3
5 ,
d
[
]
P
=
E
1
0.12
0.37
1.17
3.69
11.67
CNP P3 – depending on
800 MHz to 2.5 GHz
7
d
[
]
P
=
E
1
0.23
0.74
2.33
7.38
23.33
159