"Continuous" or "Intermittent"
In "Continuous" mode you set a constant
negative pressure for the therapy.
In "Intermittent" mode, you select a negative
pressure which alternates between two values.
In this case, you need to specify a maximum
and minimum negative pressure and set the
duration of the respective setting in minutes
in the time bar (under the pressure setting
bar respectively).
Set the desired strength
of the negative pressure in
the line with the pressure
symbol
set the time for which the
selected pressure should be
retained in the line with the
clock symbol
. In addition,
using the
keys.

4.7 Setting the sensitivity

In case of low/minimal
sensitivity
The sensitivity setting is
important for ensuring that the
message system is tailored to
the patient's personal wound
situation. The sensitivity of the
Suprasorb
set to "High", "Medium",
"Low" or "Minimal".
• However, the sensitivity must
only be set by a qualified and
authorised specialist.
• If the sensitivity is lowered
WARNING
considerably, the risk of
a blockage or leak in the
wound and tube system
going undetected and
no alarm being triggered
cannot be excluded.
• As such, if the sensitively
is lowered considerably,
particularly careful
monitoring of the therapy
must be ensured.
Setting the therapy
parameters
• Before using the
Suprasorb
CAUTION
a new patient, first check
Concerns
that the therapy parameters
qualified
set are suitable for the patient.
healthcare
professionals
CNP P3 can be
®
CNP P3 on
®
139