2.4 Filter system

There are a total of three filters integrated in the Suprasorb
the Suprasorb
CNP P3 device and two external filters in the exudate pouch.
®
The external charcoal/bacterial filter in the exudate pouch prevents oversuction. Draining stops
when the exudate pouch needs to be changed. The second filter is a ventilation filter.
The filter inside the Suprasorb
device against oversuction and contamination.
Should the internal filter become blocked, the Suprasorb
reprocessed by the L&R service centre or an L&R authorised service partner.
The activated charcoal filter in the container prevents the development of odours.

2.5 Gelling agent

The exudate drained into the exudate pouch thickens within a few minutes due to the
gelling agent contained in the exudate pouch. To avoid leaks – for example, when changing
the exudate pouch – the tube clamp on the exudate pouch tube should be closed off.
The exudate pouch incl. exudate pouch tube and tube clamp is intended
for single use only.
PLEASE NOTE

2.6 Battery

Battery, rechargeable
Charging time when battery is empty
Charging time with approx. 50% residual charge
Operating time
CNP P3 is also a charcoal/bacterial filter and protects the
®
The internal charcoal/bacterial filter is not intended for reuse and must be
replaced if it comes into contact with the exudate (blockage), once the filter
service life has elapsed or during servicing/repair work performed by L&R
or an L&R authorised service partner.
CNP P3: An internal filter inside
®
CNP P3 must be properly
®
10.8 V, 2.7 Ah – lithium ion
6-7 hours
3-3.5 hours
With battery: 10-20 hours