Error message
Status
A signal sounds.
Therapy is
interrupted.
A signal sounds.
Therapy continues
to run in the
background.
A signal sounds.
Therapy continues
to run in the
background.
A signal sounds.
Therapy continues
to run in the
background.
Possible cause
Remedy
Container is not
The container and exudate pouch may
connected.
not be connected to the device properly.
Connections open.
The cause may also be a collapsed
exudate pouch.
Or a connection between the tubes
may be open.
Please check the points above and
re-establish all connections if necessary.
Wound dressing
The tubes might not be connected
or connections
properly or the dressing itself is not
not tightly sealed.
sealed tight.
Please check both possibilities,
reconnect the tubes using
the single-piece connector or
improve the seal on the dressing.
Exudate flow
A tube clamp may have accidentally
interrupted, tube
been closed or there may be a blockage
blocked, internal
in the tubes.
filter blocked.
Please open all the tube clamps or try
to remove the blockage in the tubes.
If necessary, change the full exudate
pouch.
However, if the internal charcoal/bacterial
filter is blocked, the device needs to
be reprocessed properly by L&R or
an authorised L&R service partner!
If the message "Exudate pouch is full"
appears, it is high time to insert a new
exudate pouch. Replace the pouch in
accordance with the instructions for use
(see Section 3.4).
149