3 Preparation for commissioning

3.1 Prior to commissioning

Always keep an additional exudate pouch for the Suprasorb
obligatory for safe operation!
• Read the instructions for use through completely before you use the Suprasorb
therapy unit for the first time.
• Remove the Suprasorb
• Always position the Suprasorb
attention to the correct positioning of the device (horizontally or in the bed mount).
• All the tubes and the power supply need to be checked for damage each time before
you use the Suprasorb
connecting the tubes. Ensure that the container system and tubes are connected
correctly before switching on the device.
• Fully charge the battery prior to first use.
• Perform a function test (see Section 3.2).
WARNING
CAUTION
PLEASE NOTE
PLEASE NOTE
PLEASE NOTE
CNP P3 and its accessories from the Suprasorb
®
CNP P3 therapy unit on a solid, flat surface, paying
®
CNP P3 therapy unit. Avoiding kinking the tubes when
®
Strangulation
• Ensure that the therapy is administered in such a way that the tubes
and cables do not pose a strangulation risk.
Setting the volume
• Before using the Suprasorb
of the alarms is appropriate for the environment. Ensure that you can
hear the acoustic alarm and thus handle the error message promptly.
Electromagnetic interactions
• Medical electrical devices are subject to special safety precautions
in terms of electromagnetic compatibility and need to be installed and
commissioned in accordance with the EMC information (see Section 9)!
Incorrect accessories and spare parts
• The use of accessories and spare parts not recommended by L&R can
have a detrimental effect on the safety and functioning of the device.
• Damage caused by the use of accessories and spare parts other
than recommended or improper use of the device is not covered
by any warranty.
• Use original accessories and spare parts only.
Suctioning of other liquids
• Never suction flammable, corrosive or explosive liquids or gases.
CNP P3 available, as this is
®
CNP P3, check that the current volume
®
CNP P3
®
CNP Case.
®