6 Handling error messages

6.1 Function test

Proceed as described in Section 3.2.

6.2 Device does not start

Error
Device does not start
PLEASE NOTE
You can reach the L&R CNP service hotline 24 hours a day in the following countries:
Germany:
+49 (0)800 6647435
Austria:
+43 (0)800 808106
Switzerland
+41 (0)848 343434

6.3 Error messages on the screen and solutions

When there is an error message, the LED bar goes orange, the screen background goes
red and an acoustic alarm sounds. The volume can be changed in the settings (see Section 4
or 5 respectively). In addition, the alarm tone can be muted for two minutes using the
on the housing.
PLEASE NOTE
Possible cause
• Battery is empty
• Tube clamp is closed
• Overflow prevention system is
blocked (exudate pouch is full)
• Internal filter blocked
• Device is still in Settings mode
Contact L&R or the service partner authorised by L&R if an error occurs
that you cannot resolve with the measures described here.
Please proceed as follows:
1. When an error message appears, mute the alarm tone
2. Resolve the problem
3. Then return with
Remedy
• Connect power supply
• Check tubes are connected
correctly
• Replace exudate pouch
• Please contact the Service
department
• Confirm selection (see Section 3.2)
and start the device
; the therapy continues automatically
key