PLEASE NOTE
7 Cleaning of Suprasorb
CAUTION
CAUTION
PLEASE NOTE
Disinfection during the therapy is not expressly required when using the device on a single
patient at home.
Disinfection is generally required when the device is used in an inpatient setting.

7.1 Cleaning and disinfection of the device surfaces and the container

• Clean the device regularly and disinfect the surfaces at least once a week.
• All devices can be wiped with a damp, lint-free cloth.
When the device has been in use for an extended period, the plastic components of the housing
may exhibit slight discolouration. This does not have any effect on the functioning of the device.
If liquids come into contact with the inside of the device, the unit must be checked by L&R or
an L&R authorised service partner.
If you cannot resolve an error message successfully with
the measures outlined above, contact the L&R hotline
(see Section 11) or your CNP specialist.
®
Avoiding infections
Infectious and pathogenic microbes in the exudate can pose health risks.
• Wear appropriate disposable gloves when replacing the container.
• Replace the exudate pouch including the exudate pouch tube in
accordance with the respectively applicable hygiene regulations,
at the latest every week, before a change of patient or when full.
• Reprocessing is required whenever there is a change of patient.
Reprocessing means that all the single-use products are disposed
of and the Suprasorb
The Suprasorb
CNP P3 therapy unit must only be opened by L&R
®
or an authorised L&R service partner.
• The exudate and all components contaminated with it must be disposed
of in accordance with the facility's policies.
Cleaning the device
• Switch off the device prior to cleaning/disinfection by disconnecting
the power supply from the mains power and then from the power socket
on the Suprasorb
CNP P3.
®
• Disconnection from the power supply is only guaranteed when
the charging cable is removed from the charging port.
Use of disinfectants
• We recommend wearing appropriate disposable gloves when performing
disinfection.
• Please observe the disinfectant manufacturer's specifications.
CNP P3 therapy unit
CNP P3 therapy unit wipe-down disinfected.
®