Screen Saver And Screen Lock; Start Screen; Preset Therapy Mode; Starting The Therapy And Switching To Pause Mode - L&R Suprasorb CNP P3 Mode D'emploi

Dispositif médical d'aspiration
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.2 Screen saver and screen lock

If the screen goes dark during therapy, it can
be made visible again by simply touching it.
However, the editing function remains locked.
To make changes, unlock by pressing
until
appears.

5.4 Preset therapy mode

The screen appears with the
preset parameters for your
therapy. Here is an example.
If the screen is orange,
the therapy is paused.

5.3 Start screen

The start screen appears
for a few seconds when
the unit is switched on.
5.5 Starting the therapy and
switching to Pause mode
If bleeding occurs
The Suprasorb
designed to generate negative
pressure in the wound drainage
system and to suction exudate.
• The Suprasorb
designed to detect heavy bleeding
or pooling of blood in the wound.
• For this reason, you should not
rely on the Suprasorb
alarm system alone. The therapy
must be regularly monitored and
the exudate regularly checked for
WARNING
red discolouration so as to identify
any risk of severe deterioration in
the patient's condition.
• Should you observe bright red
blood in the tubes or exudate
pouch, the therapy must be
discontinued immediately.
Leave the wound dressing in place
at first, stem the bleeding and seek
medical assistance.
CNP P3 was
®
CNP P3 was not
®
CNP P3's
®
145

Publicité

Table des Matières
loading