3.3 Insertion of the Suprasorb
and connection of the container system
CAUTION
PLEASE NOTE
The container system of the Suprasorb
the Suprasorb
CNP Exudate Pouch with the connected exudate pouch tube and a tube clamp.
®
There are two possible combinations here:
CNP Exudate Pouch
®
Exudate pouch
• Ensure that the exudate pouch for the Suprasorb
regularly. When the exudate pouch is full, the integrated overflow prevention
system is triggered. This discontinues the therapy.
• Switch the device to Pause mode when replacing the exudate pouch.
Collapsing exudate pouch
• A leak in the container or the cover of the exudate pouch may result
in air entering the container. This, in turn, may cause the exudate pouch
to collapse.
• When assembling the container system, ensure that the side plate of
the exudate pouch is securely connected to the container.
• Ensure that all connections are secure and established correctly.
• Ensure that neither the container nor the exudate pouch is damaged.
• The separate instructions for use for the exudate pouch must also be
followed!
CNP P3 comprises the Suprasorb
®
CNP P3 is replaced
®
CNP Container,
®
A) Suprasorb
CNP
®
Exudate Pouch 250 ml
and Suprasorb
CNP
®
Container 250 ml
B) Suprasorb
CNP
®
Exudate Pouch 1,000 ml
and Suprasorb
CNP
®
Container 1,000 ml