Multiplication factor
Table 1
Guidelines and manufacturer's declaration – Electromagnetic emissions
The Suprasorb
CNP P3 therapy unit is intended for operation in an ELECTROMAGNETIC
®
ENVIRONMENT as described below.
The customer or user of the Suprasorb
in such an environment.
Measurement of
emitted interference
HF emissions
in accordance
with CISPR 11
HF emissions
in accordance
with CISPR 11
Harmonic current
emissions in accordance
with IEC 61000-3-2
Emission of voltage
fluctuations/flickers
in accordance
with IEC 61000-3-3
0 metres
500 metres
above sea level
above sea level
1.0
0.94
CNP P3 therapy unit must ensure that it is operated
®
Conformity
Group 1
Class B
Class A
Conforms
1,000 metres
above sea level
0.88
Electromagnetic environment – guidance
The Suprasorb
CNP P3 therapy unit only
®
uses HF energy for its internal workings.
As such, its HF emissions are very low and
it is unlikely that adjacent electronic devices
will be affected.
The Suprasorb
CNP P3 therapy unit is intended
®
for use in all facilities including residential buildings
and those directly connected to the mains power
grid which also supplies residential buildings.
2,000 metres
above sea level
0.78