Table 3
Guidelines and manufacturer's declaration – Electromagnetic immunity
The Suprasorb
CNP P3 therapy unit is intended for operation in an electromagnetic environment
®
as described below.
The customer or user of the Suprasorb
in such an environment.
Immunity tests
IEC 60601
test level
Conducted HF
disturbances in
accordance with
IEC 61000-4-6
3 V
Radiated HF
disturbances in
3 V/m
accordance with
80 MHz to
IEC 61000-4-3
2.5 GHz
NOTE 1: The higher frequency range applies at 80 MHz and 800 MHz.
NOTE 2: These guidelines may not be applicable in all cases. The transmission of electromagnetic parameters is influenced
by the extent to which they are absorbed and reflected by buildings, objects and people.
a In theory, it is not possible to determine the field strength of stationary transmitters, such as base stations of radio phones
and land mobile radio services, CB radio stations, AM and FM radio and TV stations in advance with accuracy. A site study
of the prevalent electromagnetic phenomena should be taken into consideration for the determination of the electromagnetic
environment with regard to stationary transmitters. If the field strength measured at the location at which the Suprasorb
therapy unit is used exceeds the above conformity levels, the device should be monitored to verify that it functions as intended.
If unusual performance characteristics are observed, additional measures such as a modified alignment or a different location
for the device may prove necessary.
b The field strength should be less than 3 V/m over the frequency range from 150 kHz to 80 MHz.
CNP P3 therapy unit must ensure that it is operated
®
Conformity
Electromagnetic environment – guidance
level
Portable and mobile wireless devices should
not be kept closer to the Suprasorb
therapy units, including all cables, than the
recommended protective distance calculated
using the applicable equation for the transmission
frequency.
Recommended protective distance:
3
d
[
=
V
3
d
[
=
3 V
E
d
[
=
E
where P is the maximum nominal capacity of
the transmitter in watts (W) in accordance with
the transmitter manufacturer's specifications
3 V/m
and d is the recommended protective distance
in metres (m). The field strength of stationary
transmitters should be less than the conformity
level at all frequencies according to a non-site
study.
devices bearing the following symbol.
5 ,
]
P
1
5 ,
]
P
80 MHz to 800 MHz
1
7
P
]
800 MHz to 2.5 GHz
1
Interference is possible in the vicinity of
b
CNP P3
®
a
CNP P3
®