Technical Data Of The Suprasorb - L&R Suprasorb CNP P3 Mode D'emploi

Dispositif médical d'aspiration
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9 Technical data of the Suprasorb

therapy unit and EMC information
Aggregate suction power
Negative pressure
Container
Power supply rated voltage
Maximum load current
Power supply line frequency
Power consumption
Battery, rechargeable
Charging time when battery is empty
Charging time with approx. 50%
residual charge
Dimensions (H x W x D)
Weight (therapy unit)
Operating time
Degree of protection in accordance
with EN 60601-1
Risk class as per 93/42/EEC,
Annex IX, Rule 11
Protection class in accordance
with EN 60601-1
CE marking
Noise emissions
Ambient conditions
REF
* Maximum negative pressure fluctuations:
The maximum value refers to sea level.
It may vary depending on altitude and atmospheric pressure. The multiplication factors listed in the table are employed
to set the maximum negative pressure based on the altitude and atmospheric pressure.
®
11 l/min
-20 to -200 mmHg*;
conversion factor: 1 kPa≈7.5 mmHg
Suprasorb
CNP Container 250 ml
®
Suprasorb
CNP Container 1,000 ml
®
100-240 V AC primary / 15 V DC secondary
1 A
50/60 Hz
15 W (charging and operation)
10.8 V, 2.7 Ah – lithium ion
6-7 hours
3-3.5 hours
175x155x70 mm with Suprasorb CNP Container
250 ml
175x260x70 mm with Suprasorb CNP Container
1,000 ml
1.15 kg with Suprasorb CNP Container 250 ml
1.38 kg with Suprasorb CNP Container 1,000 ml
Mains: Continuous operation
Battery: approx. 10-20 hours
Type BF
IIa
II
CE0123
40 dB (A)
Transport/storage: -10°C to +60°C
Operation: +5°C to +40°C
Recommended charging temperature: +15 C to +30 C
5 to 80% relative humidity, non-condensing
Atmospheric pressure: 860 hPa to 1,060 hPa
31780