Not only the pressure and time but also
the sensitivity need to be set. You can
choose between "High", "Medium",
"Low" and "Minimal" for this parameter
on the "Therapy" screen.
The sensitivity determines how sensitively
the therapy unit reacts to pressure fluctuations
and how fast a message requiring attention
appears on the screen. In the "High" setting,
this message appears particularly quickly;
in the case of "Minimal", the delay is
considerably longer.

4.8 Confirming the therapy setting

Once the setting reflects the desired therapy,
confirm with
.
You are automatically returned to the therapy
screen, where you can now see the selected
therapy settings. If you do not wish to make
any changes to the therapy settings, you can
return with
.
In this example, the
"Continuous" therapy mode
is selected, the negative
pressure set to -80 mmHg
and the sensitivity set
to "Medium".