On delivery and prior to first use of the Suprasorb
recommended to fully charge the battery and recharge after the first few uses.

3.2 Function test

Perform a function test on the charged therapy unit with the exudate pouch and container
connected before you use the Suprasorb
Proceed as follows:
1. Switch the Suprasorb
and off".
2. Start the therapy and let the therapy unit run without a connected wound dressing.
The message "System is open" should appear after 30-40 seconds.
If "System open" appears, proceed to 3.
However, should the message "System is blocked" appear, the internal filter
of the Suprasorb
®
authorised L&R service partner.
3. Confirm the "System is open" message with
4. Close the tube clamp on the exudate pouch tube. The "System is blocked" message
should appear after two minutes at the latest and be confirmed with
fails to appear even after multiple attempts, have the Suprasorb
or an authorised service partner.
If you drop the device
• Do not operate the device if the housing displays visible damage.
Contact the hotline or send the device directly to the specified address.
How to avoid damage caused by liquids
• Do not use the device in rooms with high humidity,
when bathing or showering.
• Avoid wetting the power supply, plug and screen.
• Never immerse the device in water or any other liquids
(even when switched off).
Heat accumulation and dust
• The power supply and therapy unit should not be covered during operation,
with the exception of storage in the carry bag provided for this purpose.
• Protect the therapy unit and charging cable from direct sunlight and other
sources of heat such as radiators.
• Protect the device and accessories from excessive accumulation of dust.
®
CNP P3 on as described in the Section "Switching the device on
®
CNP P3 is blocked and needs to be replaced by an
CNP P3 therapy unit, it is expressly
®
CNP P3 for therapy.
.
. If the message
CNP P3 checked by L&R
®