Section 4 Installation
This section provides necessary information to install and connect the analyzer. The installation of the analyzer
should be performed in accordance with relevant local regulations, and only by personnel trained and
experienced in the installation of the 6110 analyzer. Disconnect the power supply of the analyzer before carrying
out any work inside the analyzer. Any work inside the analyzer should be performed exclusively by personnel
specialized and authorized to work on electrical installations. In addition, and in accordance with safety standards,
it must be possible to disconnect the power supply of the analyzer in its immediate vicinity.
Electrical danger and fire hazard. Only qualified experts may perform the tasks detailed in the installation section
of this manual, while adhering to all locally valid safety regulations.
Risk of finger trapping. A silicon insert is located on the transparent front door to assist in lifting and lowering the
door when adding or removing packages for analysis. Under no circumstances must this silicon insert be
removed.
The instrument is heavy (55 kg) so extreme care must be taken when handling, to avoid damaging the instrument
or inflicting personal injury. It is highly recommended to use a pulley system (or similar) attached to the lifting
ring located on the top of the instrument when unpacking or moving it.
This is a class A product. There may be potential difficulties in ensuring electromagnetic compatibility in other
environments, due to conducted as well as radiated disturbances. In a domestic environment this product may
cause radio interference in which case the user may be required to take adequate measures.
Proper ESD (electrostatic discharge) protocols must be followed to prevent damage to the product when working
with the electronic boards.
Measurement variations of approximately 10% on CO
this variation is unlikely to occur at customer installations, user care must be taken in the routing of the instrument
power line to avoid ambient field interference.
4.1 Positioning
The instrument is heavy (55 kg) so extreme care must be taken with handling to avoid damaging the instrument or
inflicting personal injury.
Hach Lange recommends installing the instrument on a sturdy laboratory table that can more than
support the 55kg weight. Place it on a clean flat surface, and convenient to the power source. The
display screen should be at head-height for easy viewing and operation.

4.2 Pre-installation

1. Make sure you have suitable standards available before performing any calibration.
2. Make sure you have the following available:
Purge gas for O
and CO
2
Forcing gas for piercing device (Air or N
Reference gas for CO
2
6 English
sensor (CO
)
2
2
)
2
sensor (optional)
W A R N I N G
W A R N I N G
C A U T I O N
C A U T I O N
N O T I C E
N O T I C E
N O T I C E
values can occur in the 80 MHz to 1 GHz band. Although
2
C A U T I O N
6 to 7 bar (87 to 102 psia), purity > 99.9%
5.5 to 6 bar (80 to 87 psia)
5.5 to 6 bar (80 to 87 psia)