1. Product tab
Option
Description
Solubility Enter the type of beverage in the package.
Formula
Define up to five different formulae for computing a new parameter. The formula definitions must
already have been entered into the system (see
2. Package tab
Option
Comment
Brimful volume
Height
Default deformity
Deformity
coefficient
Tilt position
Pck backstop
Package type
Note: The above reference values are also displayed at the start of the measurement process for the package,
so the operator can adjust the instrument accordingly (tilt position, etc.).
3. Alarms tab
Option
Alarm parameters Select the parameter to define the alarm settings, and enter the minimum and maximum
4. Factors tab
Option
Description
Factors Small adjustments can be made to the measurements listed by entering a factor. This defaults to
1.000 for all measurements to indicate no adjustment. If this is changed then the measurement
calculated by the instrument will be multiplied by this factor to give an adjusted measurement value.
5. Options tab
Option
Headspace
Equilibrated
Slow decompress
Description
Enter a free-format alphanumeric text. This text will appear against the package
description in the main package management screen.
Enter the volume of the package when full to overflowing.
Enter the overall height of the package. This information is used if you activate the
Package size check option in
The default deformity of the package caused by internal pressure. The value is zero for
glass bottles and increases for cans and plastic bottles. This value cannot be changed.
The deformity coefficient defaults to the default deformity value. This value can be
increased or decreased if necessary.
Enter the tilt position of the package holder (optional). This should be 1, 2, or 3 and
corresponds to the number on the tilt knob on the front of the instrument .
Position of the package backstop (optional). The backstop is calibrated on a scale of
zero to seven.
Select the correct image for the type of package being measured (bottle, can, etc.). The
setups for each icon are different, so it is important to select the correct one.
Description
values allowable for each parameter listed. When the measurement value falls below
the minimum level or exceeds the maximum level for any parameter listed, an alarm will
be triggered.
Description
Check this box for a fast measurement sequence that provides data from the
headspace only.
This is only available if Headspace (previous box) is checked. Check this box if you
require the additional total O
headspace.
If checked, then after the measurement process has completed, a slow release of any
remaining pressure will take place so the package can be handled safely.
Formula management
Measurement output
on page 12.
and dissolved CO
measurements taken from the
2
2
on page 12 for details).