7. Saving, retrieving and deleting results

• The results of each successful measurement are stored
together with date and time. With more than 60 items of
measured data, the earliest items of data measured are lost.
• Select the desired user memory by pressing the M button
and then the + button.
• Press the M button again to display the average of all saved
measurements in the user memory.
• Press the M button again to display the average morn-
ing measurement for the last 7 days (morning: 5.00 a.m. –
9.00 a.m., display
) .
   
• Press the M button again to display the average even-
ing measurement for the last 7 days (evening: 6.00 p.m. –
8.00 p.m., display
) .
   
• If you continue pressing the M button, the latest individual
results are displayed with date and time.
• To clear the memory, press the M button, then the display
shows
Press the + button to select the user memory
and confirm by pressing M again. Press and hold the + and
keys simultaneously for 5 seconds (the display will show
).
• If you wish to change the user memory, see "Select memory"

8. Error messages / troubleshooting

In the event of errors, the error message E_ appears on the
display.
Error messages may appear if:
• It was not possible to record the pulse: E 1;
• You move or speak during the measurement: E2 ;
• The cuff is fastened too tightly or loosely: E3;
• Errors occur during the measurement: E4;
• The pump pressure is higher than 300 mmHg: E5;
• The batteries are almost empty
In such cases, repeat the measurement. Ensure that you do
not move or speak during the measurement. If necessary, rein-
sert or replace the batteries.

9. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy objects
on the device and cuff. Remove the batteries.

10. Technical details

Model no.
Type
Measurement
method
Measurement range
Display accuracy
22
E6.
BC 32
BC32/1
Oscillometric, non-invasive blood
pressure measurement on the wrist
Cuff pressure 0 – 300 mmHg,
systolic 50 – 250 mmHg,
diastolic 30 – 200 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown