Operation
Activate DAA mode
Touch the function button [Operating mode] (2), then
touch the [Dynamic AUTOaccurate] icon in the following
selection.
Adjust parameters
Touch the keypad of a parameter to open the keyboard.
Options:
–
Coolant temperature (3), -10 – 20 °C
–
Rotation speed (4), 10 – 280 rpm
–
Heating bath temperature (5), 20 – 80 °C)
Existing profiles with identical input values are detected
and loaded automatically!
Tap on the arrow buttons next to the actual value indi-
cator of a function to start/stop the respective process
(recirculation chiller, rotation, heating bath) individually.
Touch the [Start process] button in the footer to start all
processes at the same time: The button switches to [Stop
process]. Tap the button again to stop all processes: The
button switches to [Start process].
Touch the [Warm Up] button in the footer to acti-
vate the heating function for the heating bath
process-independently.
DAA profile
You can find the list of the available DAA profiles on
screen 2 of the DAA operating mode (to open, swipe left
once on the display).
If necessary, use the arrow buttons on the right side to
scroll through the list.
Load DAA profile and start process
Mark the required list entry and confirm with [Load] to load
the profile. The view switches back to the start window
with the operating parameters of the profile.
If necessary, you can delete profiles with [Delete].
All profile settings can be adjusted during the ongoing
process. To accept these modifications permanently in
the system, the profile has to be overwritten with the new
values after the end of the process or has to be stored as
new profile in the system. Follow the instructions on the
display.
If you adjust the intensity during the
running process, the modification must be
at least 0,5 °C. Modifications < ±0,5 °C
will not be considered.
100