Commissioning
➜
Switch off the device: Switching off the device with the flask lift in the upper end position,
the additional electronic safety function will be disabled.
➜
Switch on the device again: The device is ready for use.
Evaporation flask
CAUTION
In the event of improper handling, the evaporation flask that is to be used
may fall to the ground.
• Glass breakage!
• Risk of injury/contamination by leaking medium!
Observe the instructions in section „Mount/remove evaporation flasks" on
page 79.
Use only the evaporation flasks approved by the manufacturer.
WARNING
The smallest damage to the evaporation flask can lead to glass breakage
when pressurized.
• Glass breakage!
• Risk of injury/contamination by leaking medium!
Before each operation of the device, check the evaporation flask for
visible damage.
Mount/remove evaporation flasks
CAUTION
The heating bath is designed for a maximum operating temperature of
210°C (using silicone oil). From a liquid temperature of 50 °C, there is a
risk of injury in the event of contact.
Allow heated liquid to cool below 50 °C before removing the flask.
Use appropriate personal protective equipment (heat-resistant gloves,
eye protection, safety clothing) to remove the flask.
The evaporation flask is fixed to the vapor tube of the rotary evaporator using a special retort
clamp (EASY CLIP):
EASY CLIP open,
Flask unsecured
EASY CLIP closed,
Flask secured
79