Commissioning
Remove the heating bath from the base unit.
Empty and let cool down the heating bath.
Press the reset button [1] on the bottom of the heating bath
with a suitable tool (blunt tip, insulated or non-conductive
material).
Put the heating bath back on the base unit.
Residual heat indicator
As long as the heating bath temperature is above 50° C, the LED ring of the temperature/
vacuum control knob flashes after the end of the process. If the temperature/vacuum control
knob is assigned the vacuum control function, a temperature warning message will appear on
the display instead.
The residual heat indicator goes out when the device is switched off!
Intermediate valve (option)
Mount the optional intermediate valve [1] between the condenser [2] and the receiving flask
[3]:
Disconnect the rotary evaporator from the power supply.
Mount the supplied condenser on the rotary evaporator.
Mount the supplied [Plug-in card (intermediate valve)] in an available slot of the control
Box. Safely store the dismantled cover for later use.
Mount the intermediate valve on the condenser.
Mount a receiving flask on the intermediate valve.
Connect the connection plug of the intermediate valve to the [Plug-in card (intermediate
valve)] in the control Box.
Connection recirculation chiller or Distimatic Pro
For connecting an external recirculation chiller or an optional automatic module Hei-VOLUME
Distimatic Pro, use the RS 232 interface on the back of the control panel, see section
„Interfaces" on page 73.
For the operation of the respective unit, different settings have to be applied that are described
in detail in sections „Connectivity" on page 94 and „Remote mode" on page 108. For
more detailed information, please refer to the operating instructions supplied with the device.