Commissioning
Flask lift
Devices with manual lift
Devices with a manual lift are equipped with a lever for moving the lift up and down:
➜
Push and move the lever to the left to move the
lift up.
➜
Push and move the lever to the right to move the
lift down.
Devices with motor driven lift
For devices with motor driven lift, the lift is moved using the two lift buttons on the control
panel, see also section „Control panel" on page 73:
➜
Touch the [UP] arrow button to move the lift upwards.
➜
Touch the [DOWN] arrow button to move the lift downwards.
The movement stops as soon as you release the button!
Receiving flask
➜
Stop all running processes and the rotation of the flask.
➜
Move the flask lift to the upper end position (see section „Flask lift" on
page 81).
➜
When mounting the receiving flask, make sure that the glass cones
and sockets of the flask and the condenser do have full contact and
form a straight line (clean the glass cones/sockets before mounting the
flask!).
➜
Mount the ground clamp over the joint and tighten the ground clamp
by hand.
Heating bath
The heating bath is placed on the flexible base plate of the base unit and connected to the unit
by means of a 7-pin connecting cable. The recess provided on the bottom side ensures that the
heating bath can only be placed in the right position on the base plate.
➜
Stop all running processes and the rotation of the flask.
➜
Move the flask lift to the upper end position (see „Flask lift" on page 81).
➜
Place the heating bath on the base plate of
the basic unit: The base plate is equipped
with a guide suitable for the bottom-side
recess of the heating bath.
➜
Make sure that the heating bath is securely
fixed to the base plate and in position.
➜
Connect the supplied 7-pin connection cable
[1] between the socket of the heating bath
and the base unit.
➜
Before removing the heating bath, disconnect
the 7-pin connecting cable between the
heating bath socket and the base unit!
81