BTI ENDORET System V Manuel D'utilisation page 6

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
CAUTION: Do not centrifuge with tubes with diffe-
rent volumes. Use BTI 9 ml tubes.
Do not use the centrifuge if the rotor is not balanced. En-
sure the rotor is loaded symmetrically. Poor loading in the
rotor can damage the centrifuge.
Incorrect positioning of the tube holder of the rotor may
cause the rotor to be off balance.
Be sure to close correctly the sealing lids of the tube hol-
der before starting the centrifuged, as this may cause a
lack of balance.
INSPECTION AND CLEANING:
It is necessary to inspect and clean both the rotor and
the tube holder when necessary, so as not to accumulate
substances that may degrade the material with which it
is built.
The cleaning instructions are described in section 4.2 of
this manual.
DANGER: Inspecting the rotor, to verify the absence
of deterioration of the material, impacts, or fissures,
is important, to be able to correct any possible faults
in the unit. Remember that the worst accident that
can be anticipated would be caused by the head
breaking. Apply the same criteria to the tube holder.
OPERATING THE CENTRIFUGE:
Start-up:
Start up the centrifuge as follows:
- Connect the machine to the grid. Then press the
network filter switch, to place it in the on position.
- Touch the touchpad so that all the options remain
fixed.
Page 6
ENDORET
- To switch it on, press the ON/OFF icon, which at the
same time activates the software of the machine.
- If you wish to turn off the machine, press the ON/
OFF icon.
- To open the cover, press the open cover icon and
access the tube holder.
- Insert the tubes to centrifuge in their correspon-
ding tube holders, after unscrewing their lids, ta-
king into account that the machine must be ba-
lanced. Close the tube holder and the cover of the
centrifuge. If you have made a mistake or you have
left out a step, simply press the button to open the
cover and go back a step.
- Close the cover to be able to start up the cycle. Ve-
rify that the cover is properly closed. If it is open,
the display will show a message indicating that the
cover is open or not properly closed.
- Press the start icon to begin the centrifuge cycle.
The display will indicate the acceleration process,
until the programmed speed is reached. The time
will count down from the value selected.
- To stop it manually, press the emergency stop in-
dicator the work cycle will be automatically inte-
rrupted. The interrupted cycle message will then
appear.
- After the time selected in the work cycle has trans-
pired, the centrifuge will stop automatically, going
into the braking process.
- After the shutdown, the rotor speed will slow to
zero. Immediately afterwards, an acoustic warning
will indicate that the work cycle has ended.
- You can then remove the tubes from the chamber
by pressing the open cover icon. This opens the
cover of the centrifuge so you can access the tube
holder and unscrew the lid of each of the loaded
tube holders.
DANGER: The seal must not be handled by anybo-
dy other than BTI technical staff. Do not try to force
the door open if the lock has not yet been released.
Do not try to start a cycle by placing a weight on the
door.
OPERATIONAL PROCEDURE:
The steps to take for this technique are the following:
Drawing the blood
The patient to be treated will have a volume of between
9 and 144 ml drawn depending on the procedure to be
performed.
®
MODEL SYSTEM V