Parameters for Sterilizing the UCHR Components
The following tables provide recommended sterilization parameters for the UCHR components. Due to variations
in sterilization chambers and load configurations, it is the responsibility of the facility to determine a sterilization
protocol that ensures sterility of the device.
EtO(100% EtO)
Parameters:
Concentration:
Temperature:
Exposure Time:
Humidity:
Steam AutoClave (Pre-Vacuum)
Parameters:
Temperature:
Exposure Time:
Dry Time:
Sterrad®
Parameters:
System:
Position of Slider Screw on
Composite Post (UCHRP):
Containment
Cycle:
27 . Chapter 4 . Parameters for Sterilizing the UCHR Components
Option 1
270°F / 132°C
4 minutes
20 minutes
Option 1
100S
Lock into center of slot prior to sterilization
( see picture below)
Single self-seal sterilization pouch
Standard
Cycle 1
883 mg/L
131°F / 55°C
>= 60 minutes
>= 50% RH
Option 2
275°F / 135°C
3 minutes
16 minutes
Option 2
100NX
Any position
Single self-seal sterilization pouch
Standard
Option 1: Lock slider screw
Option 3
273°F / 134°C
18 minutes
20 minutes
in middle of post