Télécharger Imprimer la page

Integra CRW Manuel De L'utilisateur page 14

Ensemble d'anneau crânien compact universel (uchra)

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 32
EN – English
Sterile Devices (Single Use Only)
Apuzzo Stereotactic Drape
Disposable Head Ring Screws
Assembling the Universal Compact Head Ring (UCHR)
1. Attach the 4 head ring posts onto the head ring. Use the head
ring wrench to tighten the head post attachment screws.
Note
The anterior head ring posts are longer than the
posterior posts, and are angled rather than straight.
13 . Chapter 3 . Assembling the Universal Compact Head Ring
ASD1
(sterile drape)
ASD1B
(10 sterile drapes)
DHRSL5
(48 mm / box of
long screws / 5
packs, 2 screws
per pack)
DHRSS5
(34 mm / box of
short screws /5
packs, 2 screws
per pack)
Caution: Assemble the head ring posts onto the head ring prior to installing the
head ring screws. Failure to do so could potentially place increased stress on
the head ring post assembly.