Télécharger Imprimer la page

Keysight Technologies U2701A Guide D'utilisation page 55

Publicité

Table 2-6
List of Time Measurements and Voltage Measurements
Measurements Selection List
Menu
Settings
Display
Large Icon
Small Icon
List
Tile
Source
Ch 1
Ch 2
Math
Time
Wid th (Negative)
Wid th (Positive)
Frequency
Period
Rise Time
Fall Time
Duty Cycle (Positive)
Duty Cycle (Negative)
T Max
T Min
Delay
Phase Shift
Keysight U2701A/U2702A User's Guide
Comments
Change how measurement icons are displayed
Selects Channel 1, Channel 2, or Math as the waveform to be measured.
Measure the negative pulse width of a waveform (The time from the
mid-threshold of the first falling edge to the mid-threshold of the next rising
edge)
Measure the positive pulse width of a waveform (the time from the
mid-threshold of the first rising edge to the mid-threshold of the next falling
edge)
Measure the frequency of a waveform
Measure the period of a waveform (the time between the mid-threshold
crossings of two consecutive, like-polarity edge)
Measure the rise time of a waveform (the time at the upper threshold minus
the time at the lower threshold of the edge)
Measure the fall time of a waveform (the time at the lower threshold minus
the time at the upper threshold on the edge)
Measure the positive and negative duty cycle of a waveform
Duty Cycle is defined as (Pwidth / Period) * (100) which is the percentage of
the period that the waveform is high
The first time that the maximum voltage occurs on display
The first time that the minimum voltage occurs on the display
Measures the time difference from the selected edge on Channel 1 and the
selected edge on Channel 2 closest to the trigger reference point at the
middle threshold points on the waveforms
The calculated phase shift from Channel 1 to Channel 2 (in degrees)
Scope Features and Functions