Warm-up Procedures
Before performing any measurements, ensure that oscilloscope has gone through
the following warm-up procedures,
1 Connect the oscilloscope to the PC using a USB cable.
2 Power-up the oscilloscope.
3 Run the Keysight Measurement Manager.
4 Toggle channel 1 or channel 2 on.
5 Click on the Run button.
Test Considerations
For optimum performance, all procedures should comply with the following
recommendations,
– Ensure that the ambient temperature is stable.
– Ensure that the ambient relative humidity is less than 80%.
– Allow a 30 minutes warm-up period before calibration.
– Keep all the test connection cables as short as possible.
NOTE
Keysight U2701A/U2702A User's Guide
Under standalone use, users are only allowed to measure up to CAT1 
30 Vrms maximum.
For high voltage measurement up to CAT1 300 Vrms, users must install
the L-mount kit on the U2701A/U2702A. Before plugging into the
instrument chassis, ensure that the L-Mount kit installed on your modular
oscilloscope is screwed to the instrument chassis for chassis grounding
purpose (see
L-Mount Kit
10:1 probes (N2862A/N2863A) for high voltage measurements.
Installation). It is required to use the provided
Getting Started
19