Warm-up Procedures
Before performing any measurements, ensure that oscilloscope has gone through
the following warm-up procedures,
1 Connect the oscilloscope to the PC using a USB cable.
2 Power-up the oscilloscope.
3 Run the Keysight Measurement Manager.
4 Toggle channel 1 or channel 2 on.
5 Click on the Run button.
Test Considerations
For optimum performance, all procedures should comply with the following
recommendations,
– Ensure that the ambient temperature is stable.
– Ensure that the ambient relative humidity is less than 80%.
– Allow a 30 minutes warm-up period before calibration.
– Keep all the test connection cables as short as possible.
NOTE
Keysight U2701A/U2702A User's Guide
Under standalone use, users are only allowed to measure up to CAT1
30 Vrms maximum.
For high voltage measurement up to CAT1 300 Vrms, users must install
the L-mount kit on the U2701A/U2702A. Before plugging into the
instrument chassis, ensure that the L-Mount kit installed on your modular
oscilloscope is screwed to the instrument chassis for chassis grounding
purpose (see
L-Mount Kit
10:1 probes (N2862A/N2863A) for high voltage measurements.
Installation). It is required to use the provided
Getting Started