Télécharger Imprimer la page

Keysight Technologies U2701A Guide D'utilisation page 45

Publicité

Pulse Width Trigger
A pulse trigger occurs when a pulse is found in a waveform that matches the pulse
definition.
Table 2-3
Pulse Width Trigger menu
Pulse Wid th Trigger Panel Controls
Menu
Settings
Trigger
Auto
Mode
Normal
Source
Channel 1
Channel 2
External
Settings
Level
Holdoff
Polarity
Mode
> Greater than
< Less than
>< In range of
<> Out of range
To configure pulse wid th triggering
1 Select the desired trigger source on the Trigger Source panel.
2 On the Trigger Settings panel, click on the Pulse Width button.
3 Select the trigger level by adjusting the knob or by clicking on
4 Select the desired polarity on the Pulse Width Trigger Settings panel.
Keysight U2701A/U2702A User's Guide
Comments
Able to acquire waveform even when no trigger occurs
Acquires waveform only when trigger occurs
Sets Channel 1 as the trigger source
Sets Channel 2 as the trigger source
Sets External as the trigger source
Sets the voltage point on the waveform where trigger occurs
Sets the waiting period before starting a new trigger
Trigger on the positive pulse
Trigger on the negative pulse
Pulse width greater than pulse width setting
Pulse width less than pulse width setting
Pulse width within the pulse width setting range
Pulse width out of the pulse width setting range
Scope Features and Functions
2
or
.
45

Publicité

loading

Ce manuel est également adapté pour:

U2702a