Guidance and Manufacturer's Declaration - Electromagnetic Immunity
(continued):
Immunity Test
IEC 60601 Test Level
Voltage
<5 % U
dips, short
in U
) for 0.5 cycle
T
interruptions
and voltage
40 % U
variations on
in U
) for 5 cycles
T
power supply
input lines
70 % U
in U
) for 25 cycles
T
IEC 61000-4-11
<5 % U
in U
) for 5 sec
T
Power
3 A/m
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Level
(>95 % dip
<5 % U
(>95 % dip
T
T
in U
) for 0.5 cycle
T
(60 % dip
40 % U
(60 % dip
T
T
in U
) for 5 cycles
T
(30 % dip
70 % U
(30 % dip
T
T
in U
) for 25 cycles
T
(>95 % dip
<5 % U
(>95 % dip
T
T
in U
) for 5 sec
T
3 A/m
Inogen One® G2 System Specifications Chapter 8
Electromagnetic
Environment - Guidance
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the [ME EQUIPMENT or ME
SYSTEM] requires continued
operation during power
mains interruptions, it is
recommended that the
[ME EQUIPMENT or ME
SYSTEM] be powered from
an uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
43