Filter Cleaning and Replacement

The particle filter must be cleaned weekly to ensure the ease of air flow.
Remove filter from the front of the device. Clean the particle filter with a mild
liquid detergent (such as Dawn
reuse.
NOTE
To purchase additional particle filters contact your equipment provider or Inogen.
Output Filter
The output filter is intended to protect the user from inhalation of small particles in the
product gas flow. The Inogen One® G2 includes an output filter conveniently located
behind the removable cannula nozzle fitting. Inogen requires that this filter be replaced
between patients.
The output filter may be replaced by the equipment provider or by the owner using the
Output Filter Replacement Kit (RP-107).
The Inogen One G2 Concentrator must be cleaned and disinfected as per the above
instructions for each new patient. No special maintenance needs to be carried out
by the patient. Your equipment provider performs maintenance operations to assure
continued reliable service from your Inogen One G2. The manufacturer's instructions for
the preventative maintenance of the devices are defined in the service manual. All work
must be performed by trained technicians certified by the manufacturer.
DC Input Cable Fuse Replacement (For use with RP-122 & RP-222)
The Cigarette Lighter DC Power Plug contains a fuse. If the DC Input Cable is being
used with a known good power source and the power supply is not operative (no output
power and the green LED is not illuminated), the fuse may need to be replaced.
34
) and water; rinse in water and dry before
TM
It may be necessary to clean the particle filter more often in dusty
environments.