User Controls

ON / OFF Button
Press once to turn "ON"; Press and hold for one second to turn "OFF".
Audible Alarm Button
Pressing this button will toggle the Inogen One® G2's breath detection audible
alert on and off:
1. Default Mode. When the Inogen One® G2 is powered up, the breath
detection audible alert is disabled. The display's mode indication
area will show a bell icon crossed out with an X in default mode.
2. Breath Detection Alert Mode. The Inogen One® G2 will alert with
audible and visual signals for "no breath detected" when this mode is
enabled and no breath has been detected for 60 seconds. At 60 seconds,
the device will enter into auto pulse mode and once another breath is
detected, the device will exit auto pulse mode and deliver normally on
inspiration. The display's mode indication area will show a bell icon, flashing
red light and display message when the alert is enabled.
Flow Setting Control Buttons
Use the – or + flow setting control buttons to select the setting as
shown on the display. There are six settings, from 1 to 6.
Display Backlight Button
Push on; automatically turns off after 10 seconds.

User Interfaces

Display
This screen displays information regarding flow setting, power status,
battery life and errors. If you would like to change the language on the
Inogen LCD Screen contact Inogen Client Services.