Low Priority Alerts (continued)
Message Display & Text
Oxygen Low
See Manual
Remove Battery
to Cool
Comm Error
See Manual
Service Needed
02 Sensor Fail
See Manual
Condition/Action/Explanation
Concentrator is producing oxygen at a slightly low level
(<82%) for a period of 10 minutes. If condition persists,
contact your equipment provider.
Battery has exceeded its charging temperature and charging
has stopped. The battery will not charge while this alert
is present but will begin to charge when the battery
temperature returns to the normal operating range.
If battery charging is desired sooner, remove the battery
from the concentrator and allow it to cool in an open area for
approximately 10-15 minutes. Then, re-insert the battery into
the Inogen One® G2. If the problem still persists, contact your
equipment provider.
Concentrator is producing oxygen but cannot report
battery status. Replace battery. If condition persists,
contact your equipment provider.
The concentrator requires servicing at the earliest
convenience. The concentrator is operating to specification
and may continue to be used. Contact your equipment
provider to arrange for service.
The concentrator's oxygen sensor has malfunctioned.
You may continue to use the concentrator. If the condition
persists, contact your equipment provider.
Audible and Visible Signals Chapter 4