ELECTROMAGNETIC COMPATIBILITY
This CE Marked equipment has been tested and found to comply with the EMC limits for
the Medical Device Directive 93/42/EEC [EN 55011 Class B and EN 60601-1-2]. These limits
are designed to provide reasonable protection against harmful interference in a typical
medical installation.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity:
The Inogen One® G2 Oxygen Concentrator is intended for use in the electromagnetic
environment specified below. The user of the Inogen One® G2 Oxygen Concentrator
should make sure it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Compliance
Electromagnetic
Level
Environment - Guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the device, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
3 Vrms
distance:
d=1.2√
P
d=1.2√
3V/m
P
d=2.3√
P
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Inogen One® G2 System Specifications Chapter 8
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
41