Reprocessing Procedure - B.Braun Aesculap PV481 Mode D'emploi/Description Technique

Têtes de caméra 2d
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 40
Aesculap
2D camera heads PV481, PV482, PV485
Risk of injury due to misdirected leakage currents!
Before thoracic or cardiothoracic surgery, deac-
WARNING
tivate implanted defibrillators (ICD).
Before any kind of defibrillation, remove the
used camera head from the patient.
Risk of infection/injury to patients and users!
Only use the sterile cover once.
WARNING
Do not clean the sterile cover with ultrasound.
Do not reprocess the sterile cover.
Risk to patient from gas embolism!
Avoid overinsufflation (e.g. with air or inert gas)
WARNING
before high frequency surgery.
Risk to patient from damage to the optical lens!
Withdraw the endoscope slowly from the used
WARNING
trocar.
If a trocar with a manual opening valve is being
used, only withdraw the endoscope from the
trocar when the valve is open.
Risk of infection due to unsterile cover or medical
device!
Replace the sterile cover or the camera head if
WARNING
it falls on the ground or if it is touched by
unsterile objects/persons.
Malfunction in connection with magnetic reso-
nance!
Do not use the product in a magnetic resonance
CAUTION
environment.
Note
The current assignment of the buttons is displayed on the screen as "Cam-
era head information" after opening the camera menu.
Connection to camera control unit
Insert the connection plug fully into the connection socket of the cam-
era control unit until it engages in place.
10
®
6.

Reprocessing procedure

Risk of infection due to an incorrectly reprocessed
product!
Adhere to national statutory regulations,
WARNING
national and international standards and direc-
tives, and local, clinical hygiene instructions for
reprocessing.
For patients with Creutzfeldt-Jakob disease
(CJD), suspected CJD, or possible variants of
CJD, observe the relevant national regulations
concerning the reprocessing of products.
The camera head is not autoclavable. Therefore,
the camera head and corresponding sterile cover
must not be used in patients suspected of having
Creutzfeldt-Jakob disease or who have already
been diagnosed with Creutzfeldt-Jakob disease.
Mechanical reprocessing should be favored over
manual cleaning as it gives better and more reli-
able results.
Successful reprocessing of this medical device
can only be ensured through a validated repro-
cessing procedure. The operator/reprocessing
technician is responsible for this.
Risk of infection for patients and/or user due to:
Residues of cleaning and disinfectants at the
product.
WARNING
Insufficient or incorrect cleaning and disinfec-
tion of the product and accessories.
Risk of damage to and malfunction of the device
due to incorrect reprocessing!
Do not autoclave the product.
CAUTION
Risk of defects, consequential damage or a short-
ened product service life!
Follow and observe the manufacturer's require-
CAUTION
ments for reprocessing.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Aesculap pv482Aesculap pv485

Table des Matières