Technical alarm – description
Should the recorded blood pressure (systolic or diastol-
ic) lie outside the limits specified in the section "Technic-
al specifications", the technical alarm will appear on the
display indicating either " " or " Lo ". In such cases, you
should seek medical assistance and check the accuracy
of your procedure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.
NOTE:
If the unit still does not work, contact customer service at
1-800-536-0366.

11. Technical Specifications

Model no.
BM 66/U
Type
BM 66
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the up-
per arm
Measurement range Cuff pressure 0 – 300 mmHg,
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operat-
ing conditions
Permissible storage
and transport con-
ditions
Power supply
24
systolic 60– 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg, diastolic
± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard devia-
tion according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 186 mm x W 95 mm x H 56 mm
Approx. 330 g (without batteries)
22 – 42 cm / 8.7" to 16.5"
+10 °C to + 40 °C (+50 °F to + 104 °F),
≤ 90 % relative air humidity (non-
condensing)
- 20 °C to + 55 °C (-5 °F to +130 °F),
≤ 90 % relative air humidity,
800 –1050 hPa ambient pressure
4 x 1.5 V
AAA batteries