10. Technical specifications

Model No.
BM 28
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement
Cuff pressure 0-300 mmHg,
range
Systolic 50-280 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-199 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5% of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
4 x 30 memory spaces
Dimensions
L 134 mm x W 103 mm x H 60 mm
Weight
Approximately 367 g (without batteries,
with cuff)
Cuff size
22 to 42 cm
Permissible opera-
+10°C to +40°C, 15 to 85% relative
ting conditions
humidity (non-condensing)
Permissible sto-
-20°C to +55°C, 10 to 90% relative
rage conditions
humidity, 800-1050 hPa ambient pres-
sure
Power supply
4 x 1.5 V
Battery life
For approx. 300 measurements, depen-
ding on levels of blood pressure and
pump pressure
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notifica-
tion to allow for updates.
• This device complies with European Standard EN60601-
1-2 and is subject to particular precautions with regard
to electromagnetic compatibility. Please note that porta-
ble and mobile HF communication systems may interfere
with this unit. More details can be requested from the sta-
ted Customer Services address or found at the end of the
instructions for use.
• The device complies with the EU Medical Devices
Directive 93/42/EEC, the German Medical Devices Act
(Medizinproduktgesetz) and the standards EN1060-1
28