Beurer Life BM 66/U Mode D'emploi page 59

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Guidance and manufacture's declaration – electromagnetic immunity
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
a
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
BM 66/U is used exceeds the applicable RF compliance level above, the BM 66/U should be observed to verify normal operation. If abnor-
mal performance is observed, additional measures may be necessary, such as re-orienting or relocating the BM 66/U.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the BM 66/U
The BM 66/U is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user
of the BM 66/U can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF commu-
nications equipment (transmitters) and the BM 66/U as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power
of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be deter-
mined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
59
(m)
80 MHz to 800 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3
0.23
0.73
2.3
7.3
23

Publicité

Chapitres

Table des Matières
loading

Table des Matières