10 Û TROUBLESHOOTING
10.1 Û QUICK SOLUTION TO PROBLEMS
Problem
EN
No liquid outflows
from the insert during
operation
The device is on
but not working
(intervention of the APC
protection circuit)
A slight whistling sound
coming from the compact
piezo LED handpiece is
heard during operation.
26
Possible cause
The insert type does not foresee
liquid passage
The insert is obstructed
The handpiece is obstructed
The Water function is off
The insert is not correctly
tightened on the handpiece
Insert broken, worn-out, or
deformed
Handpiece or cord connector
contacts wet
Handpiece not connected to
the cord
Sync circuit malfunction
Interruption of a cord wire
Handpiece defective
The irrigation circuit has not
been completely filled
The insert is not correctly
tightened on the handpiece
Solution
Use an insert type with liquid
passage
Unscrew the insert from the
handpiece and free the insert
water passage by blowing
compressed through it.
If the problem persists, replace
the insert with a new one
Contact an Authorized Mectron
Service Center
Turn on the water function on
the dental chair
Unscrew and correctly screw the
insert again with the Mectron
torque wrench
(see paragraph 04.4)
Replace the insert with a new
one
Thoroughly dry the contacts with
compressed air
Connect the handpiece to the
cord
Contact an Authorized Mectron
Service Center
Contact an Authorized Mectron
Service Center
Contact an Authorized Mectron
Service Center
Fill the irrigation circuit
Unscrew and correctly screw the
insert again with the Mectron
torque wrench
(see paragraph 04.4)