Mectron COMPACT PIEZO LED Mode D'emploi Et D'entretien page 21

Masquer les pouces Voir aussi pour COMPACT PIEZO LED:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 89
COMPACT PIEZO LED
Lubricate with medical degree lubricant in
the indicated point;
CAUTION: do not use lubricants
based on oil or silicon.
EN
Once you've finished the cleaning
operations, check the wrench under an
adequate source of light, paying attention
to the details that could conceal dirt
residues (cavities, grooves) and, if need be,
repeat the cleaning cycle.;
Seal the wrench individually in a single-use
sterilization packet.
Sterilize the wrench in a steam autoclave.
STERILIZATION METHOD - Torque wrench
The sterilization process in steam autoclave
must be performed setting the parameters
indicated below, which are validated by
Mectron and reach the sterilization safety
levels (SAL 10
-6
- 3 times Pre-vacuum
(minimum pressure 60 mBar).
- Sterilization temperature 132°C
(interval 0°C ÷ +3°C).
- Sterilization time 4 minutes.
- Minimum drying time 10 minutes.
All the sterilization phases must be performed
by the operator in compliance to the norms
UNI EN ISO 17665-1:2007, UNI EN ISO
556-1:2002 e ANSI/AAMI ST46:2002.
20
):
6
7
8
WARNING: Checking infections - Parts
that can be sterilized - Diligently remove
all residues of organic dirt before the
sterilization.
CAUTION: Perform the sterilization
using a water steam autoclave only. Do not
use any other sterilization procedure (dry
heat, irradiation, ethylene oxide, gas, low
temperature plasma, etc.).
CAUTION: do not exceed the allowed
load of the steam sterilizer.

Publicité