Û Technical Data; Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Mectron COMPACT PIEZO LED Mode D'emploi Et D'entretien

Masquer les pouces Voir aussi pour COMPACT PIEZO LED:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 89

Guidance and manufacturer's declaration - Electromagnetic immunity

The compact piezo LED is intended for use in the electromagnetic environment specified
below. The customer or user of the compact piezo LED should ensure that it is used in such an
environment.
Immunity
test
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common mode
Voltage dips, short
<5 % U
interruptions and
in U
voltage variations
40 % U
on power supply
input lines
in U
IEC 61000-4-11
70 % U
in U
<5 % U
(>95 % dip in U
for 5 s
3 A/m
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
N.B.: U
is the AC mains voltage prior to application of the test level.
T
Test level
IEC 60601
The device continues
to work regularly and
in safety
The device continues
to work regularly and
in safety
The device continues
to work regularly and
in safety
(>95 % dip
The device can vary
T
) for 0,5 cycles
from the required
T
levels of immunity
(60 % dip
with a duration of
T
) for 5 cycles
<5% / >95% / 5 s as
T
long as the device
(30 % dip
remains in safety, no
T
) for 25 cycles
malfunctions have
T
been dected and can
be restored to
T
)
pre-test status with
T
the intervention of
the operator
The device continues
to work regularly and
in safety
09 Û TECHNICAL DATA
Compliance
level
environment - Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
commercial or hospital
environment.
Electromagnetic
EN
23

Publicité

Table des Matières
loading

Table des Matières