04.5 Û IMPORTANT INFORMATION ON THE INSERTS
WARNING:
- When the titanium nitride coating is
visibly worn out, the insert must be
replaced. Use of an overly worn out insert
reduces its efficiency.
- Diamond-coated inserts: the diamond
inserts must be replaced when the
titanium nitride coating is visibly worn
out and, at any rate, after a maximum of
10 treatments.
- Do not activate the handpiece while
the insert is in contact with the part to
be treated. Doing so, will not allow the
electronic control circuit of the device to
recognize the best point of resonance
of the insert, required for efficient and
optimum performance.
- Check the condition of wear of the
insert and that it is intact before and
during every use. If a damage or drop in
performance is noted replace it with a
new one.
- Use original Mectron inserts only. Use of
non-original inserts, in addition to
The position of the LED light on the front
4
terminal of the handpiece can be adjusted
in the following way:
• Hold the body of the handpiece and lightly
unscrew the metal ring nut located at
the base of the front terminal, rotating it
counter-clockwise.
• Rotate the front terminal so that the LED
light goes into the desired and necessary
position.
To fasten it into position, screw the metal
ring nut, rotating it clockwise.
Select the desired power and irrigation
5
levels according to the instructions given by
the chair manufacturer.
NOTE: activate the WATER function if you
need to perform a treatment that forsees
the use of water spray, otherwise switch
the WATER function to OFF.
Activate the compact piezo LED.
At the end of the treatment always put the
scaler handpiece back into its seat.
voiding the warranty, damages the
threading of the compact piezo LED
handpiece, with the risk of no longer
being able to screw the original inserts
correctly during subsequent use.
Moreover, the device settings are tested
and guaranteed to operate correctly only
when original Mectron inserts are used.
- Do not change the shape of the insert in
any way by either bending or filing it. This
could cause it to break.
- Do not use an insert that has suffered
any type of deformation.
- Do not attempt to sharpen the insert
used.
- Always check that the threaded parts
of the insert and of the handpiece
are perfectly clean – see Chapter 06 -
CLEANING AND STERILIZATION.
- If excessive pressure is applied on the
insert, it can cause the insert to break
and possibly harm the patient.
04 Û USE
EN
11