06 Û CLEANING AND STERILIZATION
3,8 bar
10 sec
Remove the insert from the ultrasonic tank
3
and rinse with distilled water;
Delicately brush the surface of the insert
4
with the brush with soft nylon bristles;
Rinse the internal canal of the insert
5
with distilled water injected under pressure
(3,8 bar) for at least 10 seconds to eliminate
all residues;
Once you've finished the cleaning
6
operations, check all the inserts under an
adequate source of light, paying attention
to the details that could conceal dirt
residues (threading, cavities, grooves) and,
if need be, repeat the cleaning cycle;
7
CAUTION: Before starting the
sterilization cycle, make sure that the
insert is thoroughly dry, both externally
and internally. For this purpose, blow
compressed air both externally and
through the internal thru hole; this will
prevent the onset of stains and haloes
on the surface or rust inside the insert.
EN
17