CLEANING AND DISINFECTION METHOD - Torque wrench
10 min
06 Û CLEANING AND STERILIZATION
Prepare an enzymatic detergent solution
1
with pH 6-9, as per the instructions of the
manufacturer;
Place the insert in the ultrasound tank
2
submerged by the enzymatic detergent
40°C
solution at 40°C, for at least 10 minutes;
Rinse with tap water to eliminate all
3
residues;
Delicately brush the surface of the wrench
4
with a brush with soft nylon bristles using
distilled water;
Dry the wrench with a clean, soft cloth
5
with low fiber-release;