09 Û TROUBLESHOOTING
09.1 Û DIAGNOSTIC SYSTEM AND SYMBOLS ON KEYBOARD
Il PIEZOSURGERY® touch is equipped with a diagnostic circuit that allows to detect operating
abnormalities and to view their type on the keyboard via their relative symbol. By using the
following chart, the user is guided toward the identification and possible solution of the
malfunction detected.
Symbols
on keyboard
Handpiece cord contacts wet
PIEZOSURGERY® touch handpiece
not connected to device
Handpiece failure
Sync circuit malfunction
Insert not tightened correctly on
handpiece
Insert broken, worn-out or deformed
Handpiece cord contacts wet
Peristaltic pump malfunction
Silicon tubing not positioned
correctly inside the pump
The device has been turned off and
on again without waiting 5 seconds
Abnormalities on electrical network
or excessive electrostatic discharges
or internal abnormalities
Turn-on procedure incorrect: the
device has been turned on with the
foot pedal pressed
Possible cause
09 Û TROUBLESHOOTING
Solution
Thoroughly dry the contacts with
compressed air
Connect the handpiece
Replace the handpiece
Contact an Authorized Mectron
Service Center
Unscrew the insert and correctly
screw it on again with the Mectron
torque wrench (see paragraph 05.4)
Replace the insert
Thoroughly dry the contacts with
compressed air
Check that there are no
impediments to pump rotation
Correctly reposition the silicon
tubing inside the pump
(see paragraph 04.3)
Turn device off and wait 5 seconds
before turning it on again
Turn device off and wait 5 seconds
before turning it on again
If the signal persists, contact an
Authorized Mectron Service Center
Check that the foot pedal is not
pressed. If the problem persists,
disconnect the pedal and, if need
be, contact an Authorized Mectron
Service Center
EN