CAUTION: To correctly use the device,
you must press the foot pedal and start
it with the insert not making contact
with the part to be treated, so that the
electronic circuit can recognize the best
point of resonance of the insert without
interferences, allowing its optimal
performance.
WARNING: Before every treatment,
make sure that the insert appropriate for
the treatment is inserted on the handpiece.
Exclusively use the Mectron torque wrench
to fasten the insert to the handpiece.
WARNING: The patient must not come
into contact with the device body or the
foot pedal.
WARNING: Do not change the insert
while the handpiece is operating, to
prevent causing wounds to the operator.
WARNING: Breakage and wear-out of
the inserts. High frequency oscillations and
wear-out may, in rare circumstances, lead
to the breakage of the insert.
Do not bend, change shape of, or
re-sharpen an insert in any way.
05.4 Û INSTRUCTIONS FOR USE
After having connected all the accessories as illustrated in Paragraph 04.3, proceed as follows:
Bending an insert or applying leverage on
it can lead to its breakage.
Deformed or otherwise damaged inserts
are susceptible to breakage during their
use. These inserts must never be used.
Excessive pressure on the inserts during
their use can lead to their breakage.
If an insert breaks, check that none of its
fragments remain in the treated part and,
at the same time, apply effective suction
to remove them.
The patient must be instructed to breathe
through his nose during the treatment,
or a dental dam must be used to prevent
the patient from ingesting fragments of
broken inserts.
When the nitride coating wears out, the
cutting efficiency decreases; re-sharpening
the insert damages it and is therefore
forbidden. Check that the insert is not
worn out. Use of a worn-out insert reduces
the cutting performance and can cause
necrosis of the bone surface treated.
During the intervention, frequently
check that the insert is intact, especially
its top.
During the intervention, avoid prolonged
contact with retractors or with metallic
instrumentation in use.