Insert the power supply cable in its
connection located at the back of the
device. Connect it to the power outlet;
EN
Equipotential plug: The device is equipped
with an additional equipotential plug
located on the rear of the console.
This plug is in accordance with DIN 42801.
Insert the connector of the equipotential
cord (optional) to the plug on the rear of the
device's console.
The purpose of additional potential
equalization is to reduce differences of
potential which can occur during operation
between the device's body and conductive
parts of other objects within the medical
environment;
Clean the touch keyboard with a soft cloth
with low fiber-release and position the
sterile protective foil on the surface of the
clean touch keyboard.
05 Û USE
05.1 Û SWITCHING THE DEVICE ON AND OFF
Turning the device on
With the device in view from its front side,
position the switch on the left of the device
body on position "I", being careful not to press
the foot pedal.
4 symbols appear on the device (ref. U inside
cover) that turn off one by one. At this point,
the device sets on the default setting and is
ready for it to be used.

Turning the device off

With the device in view from its front side,
position the switch on the left of the device
body on position "O", being careful not to
press the foot pedal.
The device turns off.
14
10
11
12
PLEASE NOTE whenever the device is
started, the following default setting
is programmed: "function" ENDO,
"irrigation" 3, "light" OFF.