TECHNICAL SPECIFICATIONS

• This Flocare
®
Infinity
pump in combination with the adapter and the Flocare
to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emissions
and immunity and are in compliance with Directive 93/42/EEC.
0344
• IEC 601: Class II
Medical Device Directive: Class IIa
• Microprocessor controlled
• BF Equipment
• Charger:
input 100-240V AC / 50-60 Hz / 0.4A Max.
output 5 V DC 2.4 A
• Battery: internal rechargeable Lithium ion battery, 3.6 V DC 2000 mAh
• Battery capacity: 24 hours at 125 ml/h
• Jet waterproof: IPX 5
• IPX 5: Splash water proof: Water jets from any direction shall have no effect on the pump.
• Dimensions: 140 x 95 x 35 mm
• Weight: approx. 392 g
• Accuracy flow rate: ±5,0% with appropriate Flocare
• Humidity:
Operation mode:
Storage:
• Temperature:
Operation mode:
Storage and transportation: -20°C to +65°C
• Atmospheric pressure:
Operation mode:
Storage and transportation: 50-106 kPa
• Occlusion detection pressure: Upstream occlusion:
Downstream occlusion:
• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies
from 0.5 to 1 ml which relates to an air bubble with an approximate length in the silicone pump segment of
6 - 13 cm.
• The Flocare
®
Infinity
can safely be operated on commercial aircraft.
• The use of other accessories, adaptors and cables than listed within this manual may result in increased
emissions or decreased immunity of the equipment of the Flocare
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in the accompanying documents.
• Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equip-
ment. If the Flocare
Infinity
pump is used adjacent to or stacked with other equipment, the pump should
®
be observed to verify normal operation.
• In case of pump scrapping, always notify your Nutricia sales unit of the pumps serial number. Pump scrap-
ping should always occur according to local legislation.
ZMNL900048_AW.indd 80-81
®
Infinity
Nurse Call are designed
®
Infinity
pump set
30% to 75% noncondensing
10% to 95 % noncondensing
+5°C to +40°C
70-106 kPa
-34 kPa (tolerance 21 kPa)
83 kPa (tolerane 21 kPa)
®
Infinity
pump.
-78-
-79-